XtremeCT
XtremeCT is specially designed to perform high-resolution pQCT studies at distal tibia and radius of humans for clinical in vivo assessment of osteoporosis.
A low X-ray dose allows regular follow-up measurements and powerful true-3D evaluation software automatically matches cortical and trabecular bone regions to previous measurements for direct comparison of density and structure characteristics.
| Density Parameters | Structural Parameters |
| Cortical Density Trabecular Density |
Cortical Thickness Trabecular Separation Trabecular Number Volume Fraction |
| Specifications | XtremeCT |
| Type | High-resolution peripheral quantitative CT No external shielding required |
| X-ray source | Sealed Air-cooled Maintenance-free 80 µm spot size 60 kVp / 40 keV (900 µA) |
| Detector | 3072 x 255 elements Fiber-optic taper 41 µm pitch |
| Resolution | ‹ 105 µm (10% MTF) 41 - 246 µm nominal isotropic (pixel size) |
| Image matrix | 512 x 512 to 3072 x 3072 pix |
| Max. scan size | 126 x 150 mm (ØxL) |
| Scan time | 2.8 min. with standard protocol (9 mm/110 slices) |
| Effective Dose | ‹ 5 µSv per measurement (ICRP91) |
| Electrical | 100 - 230 V / 50 - 60 Hz / 10A 370W |
| Dimensions | 100 - 230 V / 50 - 60 Hz / 10A 370W |
| Weight | 545 kg |
| Approvals | CE Contact us for information about other countries |